EXPERIMENT 48 · PHYSICS LABORATORY
The Dependence of Resistance on Temperature: Temperature Coefficient of Resistance of Copper
To investigate the variation of electrical resistance of a metallic copper conductor as a function of temperature, determine its temperature coefficient of resistance (α) and reference resistance at 0 °C (R₀) via linear regression, and extrapolate the inferred absolute zero of resistance. Explore a virtual apparatus, collect readings and analyse a graph.
FOLLOW THE METHOD
A reason behind every step.
To investigate the variation of electrical resistance of a metallic copper conductor as a function of temperature, determine its temperature coefficient of resistance (α) and reference resistance at 0 °C (R₀) via linear regression, and extrapolate the inferred absolute zero of resistance. Suggested independent settings: 0.0, 20.0, 40.0, 60.0, 80.0, 100.0 °C.
1Ice-Bath Calibration and Ice-Point Resistance Measurement
Immerse the copper coil sheath and calibrated mercury thermometer into a vigorously stirred equilibrium mixture of crushed distilled ice and pure water at 0.0 °C. Connect the four terminals of the digital bridge (two current leads, two potential leads) to the coil binding posts. Wait 5 minutes for complete thermal equilibrium and record the ice-point baseline resistance R₀ = 100.00 Ω.
Why this step matters
Measuring the exact ice-point resistance R₀ at 0.0 °C establishes the true zero-intercept baseline required to calculate the temperature coefficient α = m / R₀.
2Thermal Bath Assembly and Stirrer Operation
Transfer the coil and thermometer into the thermostatically controlled heating bath filled with distilled water. Activate the motorized circulation stirrer at moderate speed to ensure vigorous convective fluid mixing throughout the tank, eliminating vertical temperature stratification.
Why this step matters
Continuous circulation guarantees that the mercury thermometer bulb and the copper coil winding experience the exact same fluid temperature.
3Controlled Temperature Elevation and Thermal Stabilization
Energize the immersion heater to elevate the bath temperature to approximately 20.0 °C. Throttle the heater power and monitor the thermometer until the temperature stabilizes to within ±0.1 °C for at least 60 seconds, ensuring that the internal copper windings have reached isothermal steady state with the bath.
Why this step matters
Because copper is encapsulated within a protective metal sheath, finite thermal conduction latency requires steady-state holding time before recording resistance.
4Sequential Resistance Measurement Across Thermal Stations
Record the stabilized bath temperature T (in °C) and the corresponding four-wire resistance reading R (in Ω). Re-energize the heater to raise the bath temperature sequentially to 40.0 °C, 60.0 °C, 80.0 °C, and 100.0 °C (boiling point), recording the equilibrium resistance at each station.
Why this step matters
Sampling evenly spaced thermal stations across the entire 0 °C to 100 °C range provides a uniform leverage baseline for linear regression analysis.
5Data Tabulation and Linear Regression Fitting
Tabulate the paired observations of temperature T (horizontal axis x) and coil resistance R (vertical axis y). Compute the ordinary least-squares linear regression coefficients to determine the slope m = ΔR/ΔT (in Ω °C⁻¹) and the vertical intercept b (in Ω).
Why this step matters
Linear regression over all six temperatures averages out thermometer parallax and digital bridge quantization scatter, isolating the pure linear rate of change.
6Parameter Extraction and Inferred Absolute Zero Calculation
Calculate the temperature coefficient of resistance as α = m / b. Compute the inferred absolute zero of electrical resistance by extrapolating the best-fit line to R = 0 (T_abs = −b / m = −1 / α), and compare the experimental values against standard copper reference tables.
Why this step matters
Extrapolating to zero resistance highlights the fundamental connection between classical electron-phonon scattering and the thermal energy scale.