Cyclic Voltammetry & Diffusion lab report

A worked cyclic voltammetry & diffusion lab report with gravimetric data tables, linear regression graphs, calculations, uncertainty and conclusions.

WORKED EXAMPLE · QUANTITATIVE OBSERVATIONS

A complete example laboratory report.

Simulated demonstration data.

This report illustrates formal analytical gravimetry, volumetric calibrations, and linear regression analysis. Replace its observations with your own experimental telemetry when preparing a practical submission.

Student: __________________Student ID: __________________Course: __________________Date: __________________Partner: __________________Instructor: __________________

1. Title and aim

Diagnostic Cyclic Voltammetry and Mass Transport Kinetics via the Randles-Ševčík Equation. Determine the aqueous diffusion coefficient of the ferricyanide anion [Fe(CN)₆]³⁻ using cyclic voltammetry at a glassy carbon electrode across scan rates from 10.0 mV/s to 200.0 mV/s via the Randles-Ševčík equation.

2. Abstract

The electrochemical reversibility and mass-transport diffusion kinetics of the potassium ferricyanide redox couple, [Fe(CN)₆]³⁻ + e⁻ ⇌ [Fe(CN)₆]⁴⁻ (5.000 mM in 1.000 M KNO₃), were quantitatively investigated at 25.00 ± 0.02 °C using cyclic voltammetry at a polished glassy carbon working electrode (diameter d = 3.00 mm, geometric area A = 0.07069 cm²) versus a saturated calomel reference electrode (SCE). Voltammetric potential sweeps were recorded across six systematically varied scan rates from 10.0 mV·s⁻¹ to 200.0 mV·s⁻¹ over the potential window +0.600 V to -0.100 V vs SCE. Peak cathodic current i_pc increased monotonically from 26.18 μA at 10.0 mV·s⁻¹ to 117.06 μA at 200.0 mV·s⁻¹, while the peak-to-peak potential separation remained invariant at ΔE_p = 59.2 ± 0.4 mV and the current ratio |i_pa / i_pc| remained equal to 1.000 ± 0.005, confirming ideal Nernstian electrochemical reversibility. Linear least-squares regression of peak cathodic current i_pc against the square root of scan rate ν¹/² yielded an exceptional linear fit (i_pc = 261.75·ν¹/² + 0.005 μA) with R² = 1.000000. Application of the Randles-Ševčík equation extracted an aqueous ferricyanide diffusion coefficient D = (7.60 ± 0.11) × 10⁻⁶ cm²·s⁻¹, demonstrating near-perfect agreement with the accepted literature benchmark value (7.60 × 10⁻⁶ cm²·s⁻¹) with a percentage difference of only 0.0046%.

3. Apparatus and fixed settings

4. Theory and graph choice

Cyclic voltammetry (CV) is the cornerstone diagnostic technique of modern electroanalytical chemistry, providing quantitative access to redox potentials, electron transfer kinetics, and diffusion transport phenomena. In a cyclic voltammetric experiment, the potential of a stationary working electrode is swept linearly back and forth between two switching limits while the resulting current is recorded. For a reversible, diffusion-controlled electron-transfer reaction—such as the outer-sphere reduction of the ferricyanide anion [Fe(CN)₆]³⁻ to ferrocyanide [Fe(CN)₆]⁴⁻—the peak cathodic current i_pc scales directly with the square root of the potential scan rate ν¹/² according to the Randles-Ševčík equation. By measuring peak currents across systematic scan rates from 10.0 mV/s to 200.0 mV/s at a polished glassy carbon electrode, the aqueous diffusion coefficient D of ferricyanide is determined with sub-1% metrological precision.

i_{pc} = 0.4463 \cdot n F A c \left(\frac{n F \nu D}{R T}\right)^{1/2} \implies i_{pc} = 261.74 \cdot \nu^{1/2} \implies D = 7.60 \times 10^{-6}\text{ cm}^2\cdot\text{s}^{-1}

The primary analytical visualization presents the peak cathodic current i_pc (in μA) as a function of the square root of the potential scan rate ν¹/² (in (V·s⁻¹)¹/²). A linear Cartesian coordinate system is selected based directly on the first-principles Randles-Ševčík boundary value solution to Fick's second law for a reversible soluble-soluble redox process: i_pc = 0.4463 · n F A c · (n F ν D / R T)¹/². Under diffusion-controlled mass transport, this theoretical governing equation requires that peak current scale strictly linearly with ν¹/² and pass through the origin. Linear least-squares regression across all six experimental points reveals near-perfect linearity (R² = 1.000000) with an empirical Y-intercept b = 0.005 μA (indistinguishable from zero within experimental uncertainty). The slope m = 261.75 μA·(V·s⁻¹)⁻¹/² enables direct, unbiased extraction of the diffusion coefficient D without requiring baseline empirical corrections.

5. Method

  1. Polish the glassy carbon electrode on microcloth pads using aqueous alumina slurries (1.0 μm, 0.3 μm, and 0.05 μm successively) in figure-eight patterns. Sonicate in deionized water and ethanol for 5 minutes each to remove polishing particulates.
  2. Prepare 50.0 mL of standardized 5.000 mM K₃[Fe(CN)₆] in 1.000 M KNO₃ supporting electrolyte using degassed, triple-distilled deionized water.
  3. Assemble the three-electrode cell in the 25.00 °C jacketed vessel. Bubble high-purity argon through the solution for 15 minutes to purge dissolved oxygen, then divert argon to create a continuous inert blanket over the headspace.
  4. Configure the potentiostat between initial potential E_i = +0.600 V and switching potential E_s = -0.100 V vs SCE. Execute a blank run in 1.0 M KNO₃ to verify that residual background currents remain below 0.1 μA.
  5. Record cyclic voltammograms across scan rates ν = 10.0, 25.0, 50.0, 100.0, 150.0, and 200.0 mV/s. Record the peak cathodic current (i_pc), peak anodic current (i_pa), and peak potential separation (ΔE_p = E_pa - E_pc).
  6. Verify reversibility criteria (ΔE_p ≈ 59 mV, i_pa/i_pc ≈ 1.0). Plot peak cathodic current i_pc against ν¹/² (in (V/s)¹/²). Perform linear least-squares regression to extract the slope and compute the diffusion coefficient D using the Randles-Ševčík equation.

6. Observations and calculation table

All gravimetric observations were recorded using a calibrated 4-decimal electronic analytical balance enclosed within a draft shield chamber. Volumetric aliquots were dispensed from a Class A burette. Residuals are computed relative to the ordinary least-squares regression line.

Table 1. Seven simulated gravimetric observations; x and y represent dispensed volume and net mass metrics.
No.Voltammetric Potential Scan Rate ν (mV·s⁻¹)Peak Cathodic Current i_pc (μA)Square Root of Scan Rate ν¹/² (V¹/²·s⁻¹/²)Peak Cathodic Current i_pc (μA)Fitted y (g)Residual (g)
110.026.18000.100026.180026.1802-0.0002
225.041.39000.158141.390041.38780.0022
350.058.53000.223658.530058.5323-0.0023
4100.082.77000.316282.770082.7701-0.0001
5150.0101.38000.3873101.3800101.3804-0.0004
6200.0117.06000.4472117.0600117.05910.0009

Download the sample observations (CSV) ↓

7. Worked calculation for one observation

1. Raw Voltammetric Observables & Potential Scan Settings: Potential scan rate ν = 10.0 mV·s⁻¹ = 0.0100 V·s⁻¹. Abscissa coordinate: x = ν¹/² = √(0.0100 V·s⁻¹) = 0.1000 V¹/²·s⁻¹/². Measured peak cathodic current: y = i_pc = 26.18 μA = 2.6180e-05 A. Peak potential coordinates: E_pc = +0.182 V vs SCE, E_pa = +0.241 V vs SCE (ΔE_p = 59.2 mV). 2. Electrode Geometry & Electrolyte Stoichiometry: Working electrode: Polished glassy carbon disk (diameter d = 3.00 mm, geometric area A = 0.07069 cm²). Analyte concentration: Potassium ferricyanide c = 5.000 mM = 5.0000e-06 mol·cm⁻³ in 1.000 M KNO₃. Electron stoichiometry: Fe(CN)₆³⁻ + e⁻ ⇌ Fe(CN)₆⁴⁻ (n = 1 electron). Faraday constant F = 96,485.3 C·mol⁻¹; Gas constant R = 8.31446 J·mol⁻¹·K⁻¹; T = 298.15 K. 3. Multi-Point Randles-Ševčík Linear Least-Squares Regression: Linearized theoretical equation: i_pc = m · ν¹/² + b. Regression slope: m = 261.7436 μA·(V·s⁻¹)⁻¹/² = 2.6174e-04 A·(V·s⁻¹)⁻¹/². Regression Y-intercept: b = 0.0040 μA (consistent with zero current at zero scan rate within ±0.01 μA). Coefficient of determination: R² = 1.00000. 4. Extraction of Aqueous Ferricyanide Diffusion Coefficient D: Randles-Ševčík constant prefactor = 0.4463 · n · F · A · c · √(n F / R T) = 0.09494812 A·(cm²·s⁻¹)⁻¹/²·(V·s⁻¹)⁻¹/². Diffusion coefficient: D = [ (m · 10⁻⁶) / prefactor ]² = [ 2.6174e-04 / 0.09494812 ]² = 7.5994e-06 cm²·s⁻¹. Scaled diffusion coefficient: D = 7.60 × 10⁻⁶ cm²·s⁻¹. Literature reference benchmark at 25.00 °C: D = 7.60 × 10⁻⁶ cm²·s⁻¹ (percentage difference = 0.0079%). Transient Nernst diffusion layer thickness at ν = 10.0 mV/s: δ ≈ √(π·D·t) = 408.8 μm.

8. Graph, gradient and final calculation

12.50.142.10.271.60.3101.20.4130.70.5Square Root of Scan Rate ν¹/² (V¹/²·s⁻¹/²)Peak Cathodic Current i_pc (μA)V = 10.0 mL, m = 26.18 gV = 25.0 mL, m = 41.39 gV = 50.0 mL, m = 58.53 gV = 100.0 mL, m = 82.77 gV = 150.0 mL, m = 101.38 gV = 200.0 mL, m = 117.06 g

Circular markers represent empirical gravimetric observations. The solid line represents the ordinary least-squares linear regression fit.

x̄ = 0.2721 mL; ȳ = 71.2183 g
Σ(x − x̄)² = 0.0908 mL²
Σ(x − x̄)(y − ȳ) = 23.7778 g·mL
m = Σ(x − x̄)(y − ȳ) / Σ(x − x̄)² = 261.74792830 g/mL
b = ȳ − mx̄ = 0.00544697 g
Randles-Ševčík Fit: i_pc = 261.75·ν¹/² + 0.005 μA; Diffusion Coefficient D = 7.60 × 10⁻⁶ cm²·s⁻¹; Peak Separation ΔE_p = 59.2 mV

9. Uncertainty and comparison

Percentage difference = |estimate − theoretical value| / |theoretical value| × 100 = 0.0046%.

For n = 7 gravimetric observations, Σresidual² = 1.1107e-05; s² = Σresidual²/(n − 2). Gradient standard error = √[s²/Σ(x − x̄)²] = 0.00553. Intercept standard error = √[s²(1/n + x̄²/Σ(x − x̄)²)] = 0.00165.

Propagating the fit parameters yields an experimental density standard error of 321.04663 10⁻⁶ cm²·s⁻¹, with an ideal coefficient of determination R² = 1.000000.

In the high-precision determination of ionic diffusion coefficients via cyclic voltammetry and the Randles-Ševčík equation, metrological uncertainties and physical electroanalytical boundary conditions are rigorously quantified across electrochemical, geometric, thermostatic, and mass-transport domains: (1) Potentiostat Current Metrology & Analog-Digital Conversion: Voltammetric currents are digitized using an advanced 16-bit analog-to-digital converter with a current sensitivity range of 100 nA to 1 mA. Baseline noise is maintained below 0.05 μA. The instrument calibration traceability yields an absolute current measurement uncertainty u(i_p)/i_p = 0.08%. (2) Working Electrode Geometric Area Calibration: The glassy carbon disk electrode was calibrated by precision optical toolmaker microscopy, measuring diameter d = 3.000 ± 0.005 mm (relative uncertainty 0.17%). Because area A = πd²/4, the relative uncertainty in electrode area is u(A)/A = 2 u(d)/d = 0.33%. Mirror polishing with 0.05 μm alumina slurries achieves an electroactive roughness factor f_r = 1.02 ± 0.02. (3) Supporting Electrolyte & Deoxygenation Fidelity: The 1.000 M KNO₃ supporting electrolyte suppresses migration transference numbers below 0.4% (t_analyte < 0.004), confirming that mass transport occurs strictly by Fickian molecular diffusion. Continuous high-purity argon sparging for 15 minutes reduces dissolved oxygen background below 0.05 μA across the +0.60 V to -0.10 V vs SCE potential envelope. (4) Multi-Point Least-Squares Regression & Covariance Propagation: The Randles-Ševčík slope m = 261.744 μA·(V/s)⁻¹/² is extracted from six discrete scan rate coordinates (10.0 to 200.0 mV/s). The standard error of the slope is s_m = 0.025 μA·(V/s)⁻¹/² (relative uncertainty 0.010%), yielding an ideal coefficient of determination R² = 0.99999999. (5) Uncertainty Propagation through the Randles-Ševčík Equation: Because D is proportional to (m / A·c)², relative standard uncertainties propagate as u_r(D) = √[ 4(u(m)/m)² + 4(u(A)/A)² + 4(u(c)/c)² + (u(T)/T)² ] = √[ 4(0.0001)² + 4(0.0033)² + 4(0.0012)² + (0.00007)² ] = 0.70%. For an empirical diffusion coefficient D = 7.60 × 10⁻⁶ cm²·s⁻¹, the expanded uncertainty at the 95% confidence level (coverage factor k = 2) is U(D) = ±0.11 × 10⁻⁶ cm²·s⁻¹ (1.4%), confirming exceptional concordance with the authoritative literature benchmark (7.60 ± 0.05 × 10⁻⁶ cm²·s⁻¹).

10. Result and discussion

7.60 = 7.60 10⁻⁶ cm²·s⁻¹
CIPM standard reference value: 7.60 10⁻⁶ cm²·s⁻¹

The experimental cyclic voltammograms exhibit all four canonical diagnostic signatures of an ideal Nernstian reversible electron-transfer mechanism. First, the peak-to-peak potential separation ΔE_p = |E_pa - E_pc| = 59.2 ± 0.4 mV matches the theoretical prediction ΔE_p = 2.218 RT / nF = 59.2 mV for a one-electron transfer (n = 1) at 25.0 °C. Second, both the cathodic peak potential E_pc (+0.182 V vs SCE) and anodic peak potential E_pa (+0.241 V vs SCE) remain completely stationary across the entire 20-fold increase in scan rate from 10 mV/s to 200 mV/s, yielding an invariant formal reduction potential E°' = (E_pa + E_pc)/2 = +0.212 V vs SCE (+0.453 V vs SHE). Third, the anodic-to-cathodic peak current ratio |i_pa / i_pc| is exactly 1.000 ± 0.005, proving that the electrochemically generated ferrocyanide product is chemically stable and undergoes no coupled homogenous decomposition reactions within the voltammetric timescale. Fourth, the strictly linear scaling of i_pc with ν¹/² (R² = 1.000000) definitively proves that the reaction is controlled by semi-infinite linear molecular diffusion without kinetic limitation or surface adsorption. In unstirred solution, the application of a negative potential ramp depletes ferricyanide at the electrode surface, creating an expanding Nernst diffusion layer whose transient thickness scales as δ ≈ √(π·D·t). At faster scan rates, the electrolysis duration is shorter, restricting the diffusion layer to a much thinner zone adjacent to the electrode (δ ≈ 9.2 μm at 200 mV/s versus δ ≈ 41.2 μm at 10 mV/s). Consequently, the concentration gradient (∂c/∂x)_x=0 is much steeper at high scan rates, driving a substantially larger diffusional flux of analyte to the electrode face and generating proportionately higher peak currents. The presence of 1.000 M KNO₃ supporting electrolyte in 200-fold excess over the 5.0 mM electroactive analyte was critical in preserving pure diffusion. By providing a massive reservoir of mobile K⁺ and NO₃⁻ ions, the supporting electrolyte reduced the migration transference number of the ferricyanide anion to under 0.4%, ensuring that electrostatic attraction or repulsion by the electrode field was negligible. Furthermore, the high ionic conductivity of the solution suppressed uncompensated resistance to R_u < 25 Ω, eliminating ohmic potential distortion (iR_u < 3 mV even at the highest current of 117 μA). The extracted aqueous diffusion coefficient D = (7.60 ± 0.11) × 10⁻⁶ cm²·s⁻¹ agrees with the Stokes-Einstein relation for a hydrated octahedral complex (hydrodynamic radius r_h ≈ 0.32 nm in water at 25 °C), corroborating the rigor of modern electroanalytical metrology.

11. Precautions and limitations

['The simple Randles-Ševčík equation assumes planar semi-infinite linear diffusion; at extremely slow scan rates (< 1 mV/s), spherical edge diffusion at microelectrodes causes positive deviations.', 'At very high scan rates (> 1000 mV/s), finite heterogeneous electron transfer kinetics (quasi-reversibility) and uncompensated cell resistance (iR_u) induce peak broadening and current attenuation.', 'Alumina polishing particles can become embedded in softer electrode matrices if excessive polishing pressure is applied, altering the electroactive surface area.'] The Randles-Ševčík equation for a reversible soluble redox couple at 25.00 °C (298.15 K) is i_pc = (2.686 × 10⁵) · n³/² · A · c · D¹/² · ν¹/², where n = 1, A = 0.07069 cm² (3.00 mm diameter glassy carbon disk), and c = 5.000 × 10⁻⁶ mol·cm⁻³ (5.000 mM K₃[Fe(CN)₆] in 1.000 M KNO₃). Across scan rates from 10.0 mV/s to 200.0 mV/s, peak cathodic currents increase from 26.18 μA to 117.06 μA while peak potential separation remains strictly constant at ΔE_p = 59.2 ± 0.5 mV. Linear least-squares regression of i_pc against ν¹/² yields a slope of 261.74 μA·(V·s⁻¹)⁻¹/² with R² = 0.99999999, extracting an aqueous ferricyanide diffusion coefficient D = 7.60 × 10⁻⁶ cm²·s⁻¹, in complete agreement with authoritative electrochemical literature benchmarks.

12. Conclusion

The aqueous diffusion coefficient of the ferricyanide anion [Fe(CN)₆]³⁻ in 1.000 M KNO₃ at 25.00 °C was successfully determined as D = (7.60 ± 0.11) × 10⁻⁶ cm²·s⁻¹ using cyclic voltammetry at a polished glassy carbon electrode. The experimental peak cathodic currents scaled strictly linearly with the square root of scan rate (R² = 1.000000) across the 10.0 to 200.0 mV·s⁻¹ window, with an invariant peak potential separation ΔE_p = 59.2 mV and current ratio |i_pa / i_pc| = 1.000 confirming ideal Nernstian reversibility. The extracted diffusion coefficient agrees with authoritative electrochemistry literature benchmarks within 0.0046% relative deviation, demonstrating the high metrological precision of the Randles-Ševčík formulation for diagnostic electrochemistry.