Multi-Station Pin-on-Disk Orthopaedic Tribometer
Applies controlled reciprocating motion and normal load with continuous cycle counting.
Determine the Archard volumetric wear factor k_w of medical-grade ultra-high-molecular-weight polyethylene (UHMWPE) sliding against CoCr alloy in bovine serum. Explore interactive physiological instrumentation simulation, biomechanics, and biomedical engineering telemetry.
Ready to sample with Distance L_s = 6.0 km.
Adjust Distance L_s, then execute biomedical trial.
Biomedical engineering telemetry and physiological states are evaluated rigorously using 64-bit IEEE 754 precision. Residuals reflect ordinary least-squares regression across successive observations. Continuous data structures support high-fidelity biological and physiological modeling.
The simple pin-on-disk geometry does not replicate multi-directional cross-shear kinematics found in clinical hip gait cycles. Evaluated on medical GUR 1020 UHMWPE cylindrical pins (diameter 9.5 mm, contact area 71 mm²) articulating against mirror-polished CoCrMo disks (Ra = 0.015 μm) in 25% bovine serum at 37 °C. Constant normal load F_N = 200 N at 1.0 Hz cyclic articulation. Linear wear rate k_w = 2.1500 mm³/km.
Applies controlled reciprocating motion and normal load with continuous cycle counting.
Standardized cylindrical polymer bearing specimens machined from compression-molded rod.
Represents metallic femoral prosthetic counterface meeting ISO 5834 surface standards.
Maintains 25% bovine calf serum lubricant with 0.2% sodium azide antibacterial agent.
Measures gravimetric wear loss before and after sliding intervals with vacuum desiccator drying.
Corrects for fluid absorption mass gain to isolate true tribological wear mass loss.
Determine the Archard volumetric wear factor k_w of medical-grade ultra-high-molecular-weight polyethylene (UHMWPE) sliding against CoCr alloy in bovine serum. Standardized parameter settings: 2.0, 4.0, 6.0, 8.0, 10.0, 12.0 km.
Clean UHMWPE test pins and unloaded soak-control pins in ultrasonic baths of ethanol and deionized water, vacuum desiccate for 48 hours, and weigh on ultra-microbalance (m_0).
Measure mirror-polished CoCrMo disk surface roughness using optical profilometry, verifying Ra < 0.02 μm across the circular wear track.
Fill tribometer reservoir with 25% bovine calf serum containing 0.2% sodium azide, heat to 37.0 °C, and verify temperature stability.
Mount the UHMWPE pin in the articulated holder, apply 200 N normal force via deadweights, and initiate reciprocating circular motion at 1.0 Hz.
At sliding intervals of 2, 4, 6, 8, 10, and 12 km, remove pins, clean, desiccate, reweigh, and correct for soak fluid absorption: Δm = (m_0 - m_t) + Δm_soak.
Convert corrected mass loss to volume loss V_w = Δm / ρ (UHMWPE density ρ = 0.935 g/cm³), plot V_w versus sliding distance L_s, and calculate wear rate k_w.
Points are your recorded observations. The solid regression line validates UHMWPE Joint Wear Factor mathematical scaling.
| # | Total Sliding Distance L_s (km) | Volumetric Wear Loss V_w (mm³) | Sliding distance L_s (km) | Volumetric wear loss V_w (mm³) | Remove |
|---|
No readings yet. Execute a biomedical trial and record it.