Dual High-Fidelity Applanation Tonometers (Millar SPT-301)
Applanates surface of superficial arteries (carotid and femoral) to record non-distorted pulse waveforms.
Determine carotid-femoral pulse wave velocity (cfPWV) across varying mean arterial pressures (MAP) using dual applanation tonometry and evaluate pressure-dependent vascular stiffening. Explore interactive physiological instrumentation simulation, biomechanics, and biomedical engineering telemetry.
Ready to sample with Pressure MAP = 90.0 mmHg.
Adjust Pressure MAP, then execute biomedical trial.
Biomedical engineering telemetry and physiological states are evaluated rigorously using 64-bit IEEE 754 precision. Residuals reflect ordinary least-squares regression across successive observations. Continuous data structures support high-fidelity biological and physiological modeling.
The linear PWV-MAP relationship applies over physiological pressure ranges (70 to 120 mmHg); extreme pressures display exponential stiffening. Evaluated on a simulated human arterial tree phantom with anatomically branched elastic silicone aorta, carotid, and femoral conduits (effective carotid-to-femoral transit distance Δx = 0.450 m). Mean arterial pressure MAP controlled from 70.0 to 120.0 mmHg. Pressure-dependent PWV scaling slope k_p = 0.0450 (m/s)/mmHg with baseline PWV_0 = 2.8000 m/s (PWV = 6.85 m/s at normotensive MAP = 90 mmHg).
Applanates surface of superficial arteries (carotid and femoral) to record non-distorted pulse waveforms.
Provides anatomically scaled compliant conduits with independent mean pressure and pulse pressure control.
Measures anatomical carotid-to-femoral surface distance Δx with millimeter accuracy.
Digitizes simultaneous carotid and femoral pulse waves with sub-millisecond time resolution.
Identifies initial systolic upstroke inflection point (transit time foot) via intersecting tangent method.
Measures true transmural mean arterial pressure (MAP) to ±0.1 mmHg.
Determine carotid-femoral pulse wave velocity (cfPWV) across varying mean arterial pressures (MAP) using dual applanation tonometry and evaluate pressure-dependent vascular stiffening. Standardized parameter settings: 70.0, 80.0, 90.0, 100.0, 110.0, 120.0 mmHg.
Measure superficial surface distance from suprasternal notch to carotid site and from suprasternal notch to femoral site; calculate net aortic transit distance Δx = 0.450 m.
Position tonometer probes over the right common carotid artery and right femoral artery, adjusting hold-down pressure until maximum signal amplitude is registered.
Set circulatory simulator mean arterial pressure to 70.0 mmHg and verify stable heart rate (70 BPM) and cardiac output.
Record 10 consecutive pulse wave pairs at 2000 Hz; execute intersecting tangent algorithm to detect foot transit time delay Δt, calculating PWV = Δx / Δt.
Increment circulatory MAP through 80, 90, 100, 110, and 120 mmHg, capturing stabilized cfPWV over 10 cycles at each pressure step.
Plot cfPWV versus MAP and determine the vascular pressure sensitivity slope k_p and baseline velocity PWV_0 via linear regression.
Points are your recorded observations. The solid regression line validates Pulse Wave Velocity Arterial Stiffness mathematical scaling.
| # | Mean Arterial Pressure MAP (mmHg) | Carotid-Femoral PWV (m/s) | Mean arterial pressure MAP (mmHg) | Carotid-femoral PWV (m/s) | Remove |
|---|