Three-Op-Amp Instrumentation Amplifier Common-Mode Rejection Ratio (CMRR)

Measure differential gain and common-mode gain of a classic three-op-amp biomedical instrumentation amplifier to calculate CMRR across common-mode interference voltages. Explore interactive physiological instrumentation simulation, biomechanics, and biomedical engineering telemetry.

BIOINSTRUMENTATION · BIOSIGNALS · BIOMECHANICS · IMAGING

Your virtual biomedical engineering laboratory.

COMMON-MODE OUTPUT VOLTAGE V_OUT,CM—
PHYSIOLOGICAL & TEST CELL TELEMETRY

Ready to sample with V_cm = 1.5 V.

1.5 V
Amplifier: Medical INA118 instrumentation topology

Adjust V_cm, then execute biomedical trial.

Biomedical engineering telemetry and physiological states are evaluated rigorously using 64-bit IEEE 754 precision. Residuals reflect ordinary least-squares regression across successive observations. Continuous data structures support high-fidelity biological and physiological modeling.

Biomedical assumptions and boundary condition limits

The linear leakage model applies within the linear common-mode input range (V_cm ≤ 3.0 V). Evaluated using a low-noise medical instrumentation amplifier (INA118 precision architecture) configured for fixed differential gain A_d = 1000. Common-mode gain A_cm is evaluated across 60 Hz sinusoidal common-mode voltages from 0.5 V to 3.0 V RMS.

Biomedical apparatus & physiological sensors.

Three-Op-Amp Precision Instrumentation Amplifier

Provides buffered high-input-impedance (10 GΩ) differential biopotential amplification.

Calibrated Precision Resistor Network (0.01% matched)

Establishes difference stage symmetry to achieve high common-mode rejection.

Dual-Channel Arbitrary Function Generator

Supplies calibrated 60 Hz common-mode sinusoidal test voltages from 0.5 V to 3.0 V RMS.

Low-Noise True-RMS Digital Microvoltmeter

Measures microvolt-level common-mode leakage output V_out,cm with 100 nV resolution.

Dual Regulated DC Power Supply (±15.00 V)

Powers the operational amplifier rails with ultra-low thermal noise and ripple.

Digital Storage Oscilloscope with FFT Spectrum Analyzer

Monitors output waveform purity and verifies absence of rail saturation or clipping.

FOLLOW THE METHOD

A reason behind every step.

Measure differential gain and common-mode gain of a classic three-op-amp biomedical instrumentation amplifier to calculate CMRR across common-mode interference voltages. Standardized parameter settings: 0.5, 1.0, 1.5, 2.0, 2.5, 3.0 V.

01

Differential Gain Verification

Apply a 1.000 mV peak-to-peak 10 Hz pure differential input signal; verify that differential output reads exactly 1.000 V peak-to-peak, confirming A_d = 1000 V/V.

02

Common-Mode Test Configuration

Tie both differential non-inverting and inverting inputs together to apply identical common-mode voltage V_cm simultaneously.

03

Initial Common-Mode Excitation

Apply a 60 Hz sinusoidal common-mode voltage of 0.500 V RMS, monitoring amplifier rails to ensure linear range.

04

Microvolt Output Leakage Measurement

Measure the residual 60 Hz output voltage V_out,cm using the precision true-RMS microvoltmeter over a 30-second integration window.

05

Common-Mode Progression Sequence

Increment common-mode voltage V_cm through 1.0, 1.5, 2.0, 2.5, and 3.0 V RMS, recording V_out,cm at each stage.

06

Gain Regression and CMRR Computation

Plot V_out,cm against V_cm to determine common-mode gain A_cm = slope / 1000 (in V/V), then calculate CMRR = 20·log10(A_d / A_cm).

YOUR NOTEBOOK · 0 OBSERVATIONS

Turn measurements into biomedical engineering insight.

See a complete example report ↗

Points are your recorded observations. The solid regression line validates Instrumentation Amplifier CMRR mathematical scaling.

LINEAR REGRESSION FIT · y = mx + b
V_out,cm = A_cm · V_cm = 0.0480 · V_cm
#Common-Mode Input Voltage V_cm (V)Common-Mode Output Voltage V_out,cm (mV)Common-mode input V_cm (V)Common-mode output V_out,cm (mV)Remove

No readings yet. Execute a biomedical trial and record it.

Calculate it yourself.

CHECK YOUR UNDERSTANDING

Explain what the biomedical metrics mean.

If differential gain A_d is 1000 and common-mode gain A_cm is 0.000048 V/V, what is the CMRR in dB?

What is the primary source of 60 Hz common-mode interference on a patient during clinical ECG recording?

If common-mode input is 2.50 V and common-mode output is 0.1200 mV, what is the gain A_cm in mV/V?

What clinical circuit actively inverts and cancels common-mode body noise in medical ECG monitors?