Anechoic Hearing Aid Acoustic Test Box (Audioscan Verifit / Frye Fonix)
Provides sound-isolated anechoic acoustic test chamber with internal reference loudspeaker.
Determine the linear acoustic gain and input-output transfer slope k_g of a digital hearing aid in an ANSI S3.22 standardized 2-cc acoustic coupler across input sound pressure levels. Explore interactive physiological instrumentation simulation, biomechanics, and biomedical engineering telemetry.
Ready to sample with Input SPL_in = 60.0 dB SPL.
Adjust Input SPL_in, then execute biomedical trial.
Biomedical engineering telemetry and physiological states are evaluated rigorously using 64-bit IEEE 754 precision. Residuals reflect ordinary least-squares regression across successive observations. Continuous data structures support high-fidelity biological and physiological modeling.
The linear transfer model applies below the compression threshold; input levels > 75 dB SPL trigger non-linear WDRC compression kneepoints. Evaluated on a 16-channel digital Behind-The-Ear (BTE) hearing aid connected to a standardized ANSI S3.22 HA-2 2-cc acoustic coupler inside an anechoic sound chamber. Pure-tone acoustic test stimulus at 1.0 kHz. Linear acoustic transfer slope k_g = 0.9500 dB/dB with intercept Gain_0 = 28.5000 dB across input sound levels SPL_in from 50.0 to 75.0 dB SPL (output SPL_out from 76.00 to 99.75 dB SPL).
Provides sound-isolated anechoic acoustic test chamber with internal reference loudspeaker.
Precision metal acoustic cavity replicating human occluded ear canal volume (2.00 ± 0.02 mL).
Measures true acoustic sound pressure level inside the 2-cc coupler with NIST traceability.
Monitors input acoustic sound field at the hearing aid microphone port, maintaining closed-loop leveling.
Clinical digital hearing aid with dual omnidirectional MEMS microphones and balanced armature receiver.
Generates pure-tone test stimuli and executes automated ANSI S3.22 compliance test sequences.
Determine the linear acoustic gain and input-output transfer slope k_g of a digital hearing aid in an ANSI S3.22 standardized 2-cc acoustic coupler across input sound pressure levels. Standardized parameter settings: 50.0, 55.0, 60.0, 65.0, 70.0, 75.0 dB SPL.
Close the anechoic test box and execute automated sound field equalization to calibrate loudspeaker output across 100 Hz to 8000 Hz.
Attach the BTE hearing aid sound tube securely to the HA-2 2-cc coupler using acoustic sealing putty; seat the reference microphone within 5 mm of the hearing aid inlet.
Program hearing aid via wireless Noahlink interface into standardized ANSI S3.22 linear test mode (disable adaptive noise reduction and feedback suppression).
Set pure-tone 1.0 kHz acoustic stimulus to input level SPL_in = 50.0 dB SPL and allow 5 seconds for output stabilization.
Record the RMS sound pressure level SPL_out inside the 2-cc coupler using the calibrated laboratory condenser microphone.
Increment input sound level through 55.0, 60.0, 65.0, 70.0, and 75.0 dB SPL, recording stabilized coupler output SPL_out at each step.
Plot coupler output level SPL_out versus input level SPL_in, compute transfer slope k_g and baseline gain Gain_0 via linear regression, and calculate total acoustic gain.
Points are your recorded observations. The solid regression line validates Hearing Aid Acoustic Gain mathematical scaling.
| # | Acoustic Test Input Level SPL_in (dB SPL) | 2-cc Coupler Output Sound Level SPL_out (dB SPL) | Acoustic input sound level SPL_in (dB SPL) | 2-cc Coupler output sound level SPL_out (dB SPL) | Remove |
|---|
No readings yet. Execute a biomedical trial and record it.