Biomedical Engineering / Einthoven ECG Vector Axis
FREE TO EXPLORE
LABORATORY 09 · BIOMEDICAL ENGINEERING

Einthoven Triangle Limb Leads & Mean Electrical Axis of the Heart

Synthesize Einthoven bipolar limb leads (I, II, III), verify Einthoven's law (Lead II = Lead I + Lead III), and calculate the frontal plane mean electrical cardiac axis from QRS complex vector amplitudes. Explore interactive physiological instrumentation simulation, biomechanics, and biomedical engineering telemetry.

BIOINSTRUMENTATION · BIOSIGNALS · BIOMECHANICS · IMAGING

Your virtual biomedical engineering laboratory.

LEAD II QRS AMPLITUDE R_II—
PHYSIOLOGICAL & TEST CELL TELEMETRY

Ready to sample with Lead I R_I = 0.6 mV.

0.6 mV
ECG Bandwidth: 0.05 Hz to 150 Hz diagnostic standard

Adjust Lead I R_I, then execute biomedical trial.

Biomedical engineering telemetry and physiological states are evaluated rigorously using 64-bit IEEE 754 precision. Residuals reflect ordinary least-squares regression across successive observations. Continuous data structures support high-fidelity biological and physiological modeling.

Biomedical assumptions and boundary condition limits

The linear dipole model assumes an electrically homogeneous, isotropic volume conductor. Evaluated on an anatomical conductive torso phantom with calibrated dipole orientation (normal mean electrical axis θ = 51.3°). Linearized projection ratio between Lead I and Lead II is evaluated across QRS amplitudes from 0.2 mV to 1.2 mV.

Biomedical apparatus & physiological sensors.

Multi-Lead Diagnostic Clinical ECG Recorder

Provides 12-lead acquisition with 0.05 to 150 Hz diagnostic clinical frequency bandwidth.

Anatomical Electrolytic Torso Vector Phantom

Saline-filled humanoid torso model featuring rotating central dipole generator.

Programmable Cardiac Waveform Synthesizer

Generates standard P-QRS-T waveforms with calibrated dipole amplitudes and heart rates.

Four-Limb Sintered Ag/AgCl Strap Electrodes

Captures RA, LA, LL, and RL biopotentials with < 2.0 kΩ contact impedance.

Precision Active Right-Leg Drive Circuit

Suppresses 60 Hz common-mode mains interference by 40 dB via negative feedback.

Biomedical Telemetry Vector Display Software

Computes instantaneous frontal plane vector loops and calculates numerical axis angle.

FOLLOW THE METHOD

A reason behind every step.

Synthesize Einthoven bipolar limb leads (I, II, III), verify Einthoven's law (Lead II = Lead I + Lead III), and calculate the frontal plane mean electrical cardiac axis from QRS complex vector amplitudes. Standardized parameter settings: 0.2, 0.4, 0.6, 0.8, 1.0, 1.2 mV.

01

Limb Lead Electrode Placement

Affix Ag/AgCl electrodes to standard phantom limb terminals: RA (Right Arm), LA (Left Arm), LL (Left Leg), and RL (Right Leg ground reference).

02

Phantom Dipole Axis Standardization

Set the internal dipole generator angle to standard anatomical normal orientation of +51.3° in the frontal plane.

03

Einthoven's Law Verification

Record baseline resting traces; verify that instantaneous voltages satisfy Einthoven's law: V_II(t) = V_I(t) + V_III(t) at all points along the QRS.

04

Initial QRS Amplitude Setting

Set cardiac synthesizer dipole strength to produce a Lead I QRS peak amplitude R_I of 0.20 mV; log stabilized Lead II amplitude R_II in millivolts.

05

Dipole Amplitude Progression

Increment dipole strength so that Lead I reads 0.40, 0.60, 0.80, 1.00, and 1.20 mV, recording corresponding Lead II amplitudes.

06

Vector Regression and Axis Calculation

Plot R_II versus R_I to determine vector ratio k_axis, calculating frontal plane axis angle θ = arctan[(2·R_II - R_I) / (√3 · R_I)].

YOUR NOTEBOOK · 0 OBSERVATIONS

Turn measurements into biomedical engineering insight.

See a complete example report ↗

Points are your recorded observations. The solid regression line validates Einthoven ECG Vector Axis mathematical scaling.

LINEAR REGRESSION FIT · y = mx + b
R_II = k_axis · R_I + R_0 = 1.2500 · R_I + 0.1500
#Lead I QRS Amplitude R_I (mV)Lead II QRS Amplitude R_II (mV)Lead I QRS amplitude R_I (mV)Lead II QRS amplitude R_II (mV)Remove

No readings yet. Execute a biomedical trial and record it.

Calculate it yourself.

CHECK YOUR UNDERSTANDING

Explain what the biomedical metrics mean.

According to Einthoven's triangle law, which equation is always strictly satisfied?

What is considered the normal range for the frontal plane mean electrical cardiac axis in an adult?

If Lead I amplitude is 0.80 mV and follows R_II = 1.2500 · R_I + 0.1500, what is the Lead II amplitude?

What is Wilson's Central Terminal (WCT) used for in clinical 12-lead ECG machines?