Precision Controlled-Stress Cone-and-Plate Rheometer
Applies controlled angular velocity and measures torque with micro-Newton-meter magnetic bearing.
Determine the Casson asymptotic plastic viscosity μ_c and yield stress τ_y of whole human blood across low-to-medium shear rates using cone-and-plate rheometry. Explore interactive physiological instrumentation simulation, biomechanics, and biomedical engineering telemetry.
Ready to sample with Shear Rate γ̇ = 64.0 s⁻¹.
Adjust Shear Rate γ̇, then execute biomedical trial.
Biomedical engineering telemetry and physiological states are evaluated rigorously using 64-bit IEEE 754 precision. Residuals reflect ordinary least-squares regression across successive observations. Continuous data structures support high-fidelity biological and physiological modeling.
The Casson model is an empirical continuum approximation; it applies across 10 to 300 s⁻¹ but does not capture zero-shear viscoelasticity. Evaluated on anti-coagulated human whole blood (Hct = 45%, EDTA treated) at 37.0 °C in a titanium cone-and-plate rheometer (cone angle θ = 1.0°, diameter 60 mm). Standard shear rates γ̇ = 16, 36, 64, 100, 144, 196 s⁻¹ correspond to √γ̇ = 4, 6, 8, 10, 12, 14 s⁻⁰·⁵. Casson slope √μ_c = 0.0620 Pa⁰·⁵·s⁰·⁵ (asymptotic viscosity μ_c = 0.00384 Pa·s = 3.84 cP) with yield root √τ_y = 0.1800 Pa⁰·⁵ (yield stress τ_y = 0.0324 Pa).
Applies controlled angular velocity and measures torque with micro-Newton-meter magnetic bearing.
Generates strictly uniform shear rate γ̇ = ω / tan(θ) throughout the entire measurement fluid volume.
Maintains physiological temperature preventing blood viscosity thermal artifacts.
Prevents plasma water evaporation and blood drying during multi-point shear sweeps.
Maintains uniform erythrocyte resuspension prior to pipetting into the rheometer gap.
Delivers exact sample volume (0.50 mL) to perfectly fill the cone-plate truncation gap.
Determine the Casson asymptotic plastic viscosity μ_c and yield stress τ_y of whole human blood across low-to-medium shear rates using cone-and-plate rheometry. Standardized parameter settings: 16.0, 36.0, 64.0, 100.0, 144.0, 196.0 s⁻¹.
Mix EDTA-anticoagulated whole blood on roller mixer for 10 minutes; verify standard hematocrit (Hct = 45 ± 1%) via microhematocrit centrifuge.
Set rheometer Peltier bottom plate to exactly 37.0 °C and allow 15 minutes for thermal equilibrium.
Zero the normal force and calibrate the mechanical contact gap; pipette exactly 500 μL of blood onto the plate center and lower cone to truncation height (50 μm).
Engage the humidified solvent trap and apply a 30-second preshear at 300 s⁻¹ to completely disperse pre-existing rouleaux aggregates.
Step through shear rates γ̇ = 16, 36, 64, 100, 144, and 196 s⁻¹, holding each step for 15 seconds to achieve steady torque before recording shear stress τ.
Compute √γ̇ and √τ, plot √τ versus √γ̇, and extract Casson asymptotic viscosity root √μ_c from the slope and yield stress root √τ_y from the y-intercept.
Points are your recorded observations. The solid regression line validates Blood Rheology Casson Model mathematical scaling.
| # | Fluid Shear Rate γ̇ (s⁻¹) | Measured Shear Stress τ (Pa) | Square root of shear rate √γ̇ (s⁻⁰·⁵) | Square root of shear stress √τ (Pa⁰·⁵) | Remove |
|---|
No readings yet. Execute a biomedical trial and record it.