Biomedical Engineering / FES Biphasic Charge Balance
FREE TO EXPLORE
LABORATORY 41 · BIOMEDICAL ENGINEERING

Biphasic Functional Electrical Stimulation (FES) Charge Balance & Injection

Verify net zero electrochemical charge balance and evaluate cathodic phase charge injection Q_p as a function of stimulus pulse current I_stim in a Functional Electrical Stimulation (FES) neuroprosthesis. Explore interactive physiological instrumentation simulation, biomechanics, and biomedical engineering telemetry.

BIOINSTRUMENTATION · BIOSIGNALS · BIOMECHANICS · IMAGING

Your virtual biomedical engineering laboratory.

CATHODIC PHASE INJECTED CHARGE Q_P—
PHYSIOLOGICAL & TEST CELL TELEMETRY

Ready to sample with Current I_stim = 15.0 mA.

15.0 mA
Stimulator: Constant-current biphasic FES unit (35 Hz, symmetric biphasic)

Adjust Current I_stim, then execute biomedical trial.

Biomedical engineering telemetry and physiological states are evaluated rigorously using 64-bit IEEE 754 precision. Residuals reflect ordinary least-squares regression across successive observations. Continuous data structures support high-fidelity biological and physiological modeling.

Biomedical assumptions and boundary condition limits

The simple Q = I · t relationship assumes a perfect rectangular current pulse; real inductive leads introduce finite rise and fall times. Evaluated on a clinical constant-current FES neuroprosthesis (Bioness NESS L300 architecture) delivering charge-balanced symmetric biphasic rectangular pulses (pulse duration t_pw = 250 μs = 0.2500 ms, frequency 35 Hz) into a 1.0 kΩ parallel RC tissue load in 37.0 °C saline. Cathodic injected charge slope t_pw = 0.2500 μC/mA across pulse currents I_stim from 5.0 to 30.0 mA.

Biomedical apparatus & physiological sensors.

Clinical Multichannel Constant-Current FES Stimulator

Delivers programmable charge-balanced biphasic current pulses from 1 to 50 mA with 250 μs duration.

Platinum-Iridium Neurostimulation Electrode Pair (10 mm²)

Biocompatible stimulation electrodes with high capacitive charge injection capacity (Q_inj).

Electrochemical Tissue Impedance Load Phantom (1.0 kΩ || 10 nF)

Accurately models transcutaneous skin-electrode interface impedance and double-layer capacitance.

Precision Current-Sense Resistor & Analog Integrator Circuit

Integrates current waveforms in real time to measure injected phase charge Q_p with 0.01 μC accuracy.

High-Resolution Dual-Channel Digital Oscilloscope

Captures simultaneous cathodic and anodic phase waveforms and monitors residual DC charge balance.

Thermostatically Controlled Physiological Saline Bath (37.0 °C)

Maintains 0.9% NaCl electrolyte environment for electrochemical electrode testing.

FOLLOW THE METHOD

A reason behind every step.

Verify net zero electrochemical charge balance and evaluate cathodic phase charge injection Q_p as a function of stimulus pulse current I_stim in a Functional Electrical Stimulation (FES) neuroprosthesis. Standardized parameter settings: 5.0, 10.0, 15.0, 20.0, 25.0, 30.0 mA.

01

Tissue Phantom Load Connection and Calibration

Connect the FES stimulation output across the 1.0 kΩ precision non-inductive load in parallel with 10 nF double-layer capacitor; verify zero DC bias voltage.

02

Biphasic Waveform Symmetry Configuration

Program stimulator to deliver charge-balanced symmetric biphasic pulses: pulse width t_pw = 250 μs cathodic, 50 μs interphase delay, followed by 250 μs anodic reversal at 35 Hz.

03

Initial Current Step Initialization

Set stimulation current amplitude to I_stim = 5.0 mA and initiate continuous stimulation train.

04

Cathodic Phase Charge Integration

Record the voltage drop across the precision 10.0 Ω sensing resistor, integrate current over the 250 μs cathodic duration, and calculate injected charge Q_p = I_stim · t_pw in micro-Coulombs.

05

Net DC Charge Balance Verification

Integrate the complete cycle (cathodic plus anodic phases) over 100 pulses to confirm net DC offset charge is strictly < 0.01 μC (safe charge balance).

06

Current Progression Sequence

Increment stimulation current through 10.0, 15.0, 20.0, 25.0, and 30.0 mA, recording cathodic injected charge Q_p at each current step.

07

Charge Injection Linear Regression Analysis

Plot injected charge Q_p versus stimulation current I_stim and verify that the regression slope matches the programmed pulse duration t_pw = 0.2500 μC/mA.

YOUR NOTEBOOK · 0 OBSERVATIONS

Turn measurements into biomedical engineering insight.

See a complete example report ↗

Points are your recorded observations. The solid regression line validates FES Biphasic Charge Balance mathematical scaling.

LINEAR REGRESSION FIT · y = mx + b
Q_p = t_pw · I_stim = 0.2500 · I_stim (where t_pw = 250 μs = 0.2500 ms)
#Stimulation Pulse Current I_stim (mA)Cathodic Phase Injected Charge Q_p (μC)Stimulation pulse current I_stim (mA)Cathodic injected charge Q_p (μC)Remove

No readings yet. Execute a biomedical trial and record it.

Calculate it yourself.

CHECK YOUR UNDERSTANDING

Explain what the biomedical metrics mean.

Why must neurostimulation pulses be strictly charge-balanced biphasic waveforms?

What is the injected electrical charge Q_p for a 20.0 mA pulse lasting 250 μs (0.250 ms)?

Which polarity of electrical pulse is most effective at initiating an action potential in an adjacent nerve axon?

What does the Shannon criterion define in clinical neuroprosthetics?