Screen-Printed GOx Platinum Biosensor Strip
Working electrode with immobilized glucose oxidase and permselective polyurethane outer membrane.
Evaluate the steady-state catalytic current of a glucose oxidase (GOx) platinum-carbon amperometric electrode across blood glucose concentrations to evaluate sensor sensitivity and linear dynamic range. Explore interactive physiological instrumentation simulation, biomechanics, and biomedical engineering telemetry.
Ready to sample with Glucose C = 6.0 mmol/L.
Adjust Glucose C, then execute biomedical trial.
Biomedical engineering telemetry and physiological states are evaluated rigorously using 64-bit IEEE 754 precision. Residuals reflect ordinary least-squares regression across successive observations. Continuous data structures support high-fidelity biological and physiological modeling.
The linear kinetic model applies across the clinical range (2.0 to 12.0 mmol/L); higher levels enter enzyme saturation. Evaluated using a screen-printed three-electrode sensor (platinum working electrode, Ag/AgCl reference, carbon counter) polarized at +0.650 V in PBS buffer (pH 7.4, 37.0 °C). Evaluated across clinical normo- and hyperglycemic concentrations (2.0 to 12.0 mmol/L).
Working electrode with immobilized glucose oxidase and permselective polyurethane outer membrane.
Applies constant +0.650 V anodic polarizing potential and records microampere currents.
Maintains reaction buffer at 37.00 ± 0.05 °C with constant 250 RPM magnetic stirring.
Dispenses calibrated standard glucose stock solution aliquots into the reaction cell.
Traceable 100 mmol/L β-D-glucose stock solution in saturated benzoic acid preservative.
Digitizes anodic oxidation current at 10 Hz with 1.0 nA measurement precision.
Evaluate the steady-state catalytic current of a glucose oxidase (GOx) platinum-carbon amperometric electrode across blood glucose concentrations to evaluate sensor sensitivity and linear dynamic range. Standardized parameter settings: 2.0, 4.0, 6.0, 8.0, 10.0, 12.0 mmol/L.
Fill the glass cell with 10.0 mL of phosphate-buffered saline (PBS, pH 7.4) at 37.0 °C; insert the three-electrode sensor strip and start magnetic stirring at 250 RPM.
Apply +0.650 V polarizing potential to the platinum working electrode; wait 10 minutes until background capacitive current stabilizes (< 0.05 μA).
Inject calibrated volume of glucose stock to achieve 2.0 mmol/L (36 mg/dL); allow 30 seconds for enzymatic steady-state current plateau.
Record steady-state anodic oxidation current I_a in microamperes over a 15-second integration window.
Perform sequential spiked standard additions to achieve 4.0, 6.0, 8.0, 10.0, and 12.0 mmol/L (up to severe hyperglycemia ~216 mg/dL), logging I_a at each step.
Plot catalytic current I_a against glucose concentration C_glu, performing linear regression through zero to determine biosensor sensitivity S_glu.
Points are your recorded observations. The solid regression line validates Glucose Biosensor Amperometry mathematical scaling.
| # | Glucose Concentration C_glu (mmol/L) | Catalytic Oxidation Current I_a (μA) | Glucose concentration C_glu (mmol/L) | Catalytic current I_a (μA) | Remove |
|---|
No readings yet. Execute a biomedical trial and record it.